Development and validation of an automated solid-phase extraction-LC-MS/MS method for the bioanalysis of fluoxetine in human plasma


Ishtiaque Ahmad1, Zabih Ullah2, Mohd Ibrahim Khan3, Abdullah Khaloofa Alahmar4, Mohd Faiyaz Khan4 
1 Scientific Research Center, Prince Sultan Military Medical City, Al-Kharj, Riyadh, Saudi Arabia
2 Department of Pharmaceutical Sciences, College of Pharmacy and Dentistry, Buraydah Colleges, Alqassim, Saudi Arabia
3 Department of Pharmaceutical, Sciences SIHAS, Sam Higginbottom, University of Agriculture Technology and Sciences, Naini, Prayagraj, India
4 Department of Clinical Pharmacy, College of Pharmacy, Prince Sattam Bin Abdulaziz University, Al-Kharj, Riyadh, Saudi Arabia

Correspondence Address:
Dr. Mohd Faiyaz Khan
Department of Clinical Pharmacy, College of Pharmacy, Prince Sattam Bin Abdulaziz University, Al-Kharj, Riyadh
Saudi Arabia

A wide-range, specific, and precise liquid chromatography tandem mass spectrometric (LC-MS/MS) technique for quantifying fluoxetine (FLX) in human plasma was developed using the RapidTrace® automated solid-phase extraction (SPE) method; the analyte and internal standard (IS) were extricated on Oasis MCX SPE cartridges. Acetonitrile and 5 mM ammonium formate buffer (90:10 v/v) were used as mobile phase to achieve chromatographic separation on the reverse phase (C<sub>18</sub> column). The analyte and IS were ionized using +ve electrospray ionization approach which was further traced by multiple-reaction monitoring on a tandem mass spectrometer. To quantify the FLX and FLX-d5, the parent-to-daughter ion transition of m/z of 310.0/44.1 and 315.0/44.0 was used, respectively. The method demonstrated a linear active limit of 0.20–30 ng/ml with recoveries ranging from 63.04% to 79.39% for quality control samples and 61.25% for IS samples. The concentrations over the calibration range demonstrated acceptable precision and accuracy. Due to the high inconsistency of the FLX concentration data, the minimum threshold of the assay was kept at 0.20 ng/ml. The flow rate was maintained at 500 <Symbol>μ</Symbol>L/min, and the time for sample analysis for each injection was 3.5 min. The method was found to be specific, sensitive, and faster with minimum utilization of organic solvents and was utilized further for metabolic and pharmacokinetic studies.


How to cite this article:
Ahmad I, Ullah Z, Khan MI, Alahmar AK, Khan MF. Development and validation of an automated solid-phase extraction-LC-MS/MS method for the bioanalysis of fluoxetine in human plasma.J Adv Pharm Technol Res 2021;12:267-273


How to cite this URL:
Ahmad I, Ullah Z, Khan MI, Alahmar AK, Khan MF. Development and validation of an automated solid-phase extraction-LC-MS/MS method for the bioanalysis of fluoxetine in human plasma. J Adv Pharm Technol Res [serial online] 2021 [cited 2022 Jan 18 ];12:267-273
Available from: https://www.japtr.org/article.asp?issn=2231-4040;year=2021;volume=12;issue=3;spage=267;epage=273;aulast=Ahmad;type=0